Test fixtures for a camera head, coupler and camera control unit: laser intensity, interface reliability, focus ring torque, and front cover vibration testing.
- R&D Mechanical Engineering Co-op
- Stryker Endoscopy
- January 2026 – June 2026
- SOLIDWORKS, Creo, MJF + MJP printing, Instron, Torsional Instron, Vibration table, Manual mill + lathe
- R&D Engineering · Design for Test · Mechanical Reliability · Verification + Validation · Medical Devices

Specifications
- MJF, MJP for alignment-critical features
- Optical, torque, vibration, lifecycle
- Camera head, coupler, camera control unit
Laser intensity fixture
A laser intensity testing setup, with and without a scope in the path, at variable distance.
- Rev 1 — basic centering, discrete distance markings.
- Rev 2 — changed clearances, stronger base.
- Rev 3 — added insert system and y-axis lock.
- Rev 4 — switched critical alignment features to MJP.
- Final — complete locking system, constrained through the coupler instead of the scope, coupler-only version added.






Focus ring fixture
A focus ring geometry change meant actuation torque had to be tested across the lifecycle, across different fin styles.
I used this fixture to perform torsional Instron testing and torque gauge testing.





Interface mechanical reliability
Simulates scope use with a coupler, and fully constrains the next-generation coupler to the test fixture.




Camera control unit — front cover and vibration
Front cover mounting and cable receptacle changes needed cable receptacle backout resilience verified under vibration.



Proprietary information withheld.
Next project
Sageware Textile Upcycling SystemLet’s talk.
Open to co-op, internship and project conversations in mechanical design, product development and R&D.